The patient was undergoing a coil embolization procedure using a penumbra coil 400 coils.During the procedure, the physician repositioned the pc400 coil multiple times.The physician attempted to advance the pc400 coil again, yet it would not advance.The pc400 removed and the procedure continued using additional coils.There was no report of an adverse effect on the patient.
|
Result: the pc400 coil was intact with the pusher assembly.The pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked approximately 5.0 cm from the proximal end.The pusher assembly also had multiple kinks along the entire hypo-tube.The coil was crushed approximately 5.0 cm from the proximal end.The outer diameter of the coil (in the undamaged region), and the distal detachment tip (ddt) were measured and determined to be within specification.Conclusion: the complaint has been evaluated.The complaint indicates that the pc400 coil could be withdrawn, but could not be advanced.Evaluation of the returned product revealed that the coil was damaged in several locations.In addition, the coil pusher assembly was kinked.Based on the condition of the returned device, it appears that the coil was damaged during manipulation in the patient causing resistance when attempting to advance the coil during placement.The multiple kinks along the length of the pusher assembly were likely due to handling the device during packaging for return.These devices are 100 % functionally tested and visually inspected for damage during process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|