Catalog Number 000000000000080300 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/04/2014 |
Event Type
malfunction
|
Event Description
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: (b)(6) the disposable set is unavailable for return because the customer discarded it.This report is being filed due to a device malfunction that has the potential for injury.
|
|
Manufacturer Narrative
|
Investigation: the disposable set was unavailable for return and investigation.The run data file (rdf) was analyzed for this event.The device history records (dhr) were reviewed for this event.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Root cause: the signals in the run data file indicate that the higher than expected wbc content in the platelet product was likely a result of an escape of wbcs from the lrs chamber toward the end of the procedure.Based on the available information, it cannot be ruled out that the higher than expected wbc content in the platelet product could be donor related.
|
|
Search Alerts/Recalls
|