Catalog Number 000000000000080400
Device Problem High Test Results (2457)
Patient Problem No Patient Involvement (2645)
Event Date 12/03/2014
Event Type  malfunction  
Event Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: (b)(4).The disposable set is unavailable for return because the customer discarded it.This report is being filed due to a device malfunction that has the potential for injury.
 
Manufacturer Narrative
Investigation: the disposable set was unavailable for return and investigation.The run data file (rdf) was analyzed for this event.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Root cause: the analysis of the run data file did not find a conclusive cause for the higher than expected wbc content in the platelet product reported for this collection.No unusual process variable was identified and the signals in the run data file indicate that the trima accel system operated as intended.Based on the available information, it cannot be ruled out that there ported results could be donor related.It also cannot be ruled out that a sampling, calculation,or other process error could have contributed to the higher than expected wbc content in the platelet product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL PLATELET, SAMPLER, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
jessica kim
10811 w. collins ave
lakewood, CO 80215
3032314812
MDR Report Key4357480
MDR Text Key15112797
Report Number1722028-2014-00511
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK140158
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/01/2016
Device Catalogue Number000000000000080400
Device Lot Number10W3116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/08/2014
Initial Date FDA Received12/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/15/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-