Result: the 5max ace reperfusion catheter was fractured approximately 1.0 cm from the hub, underneath the strain relief.The 5max ace reperfusion catheter was kinked approximately 60.5 cm from the distal tip.Conclusion: the complaint has been evaluated.The complaint indicated a fracture in the proximal shaft of the 5max ace reperfusion catheter.Evaluation of the returned product confirmed the complaint.This type of damage typically occurs when the device is improperly handled during removal from the packaging or during preparation for use.If the catheter is removed from the packaging hoop or manipulated during insertion into the patient at an angle, the shaft may fracture due to excess force and bending of the catheter.These devices are 100% visually inspected for damage during process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|