• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. MALLEABLE SUCTION MEDIUM, STANDARD TIP; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. MALLEABLE SUCTION MEDIUM, STANDARD TIP; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735016
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/29/2014
Event Type  malfunction  
Event Description
A medtronic representative reported that a malleable suction was unable to be navigated during functional endoscopic sinus surgery (fess) procedure after it was able to be navigated previously during the surgery.The 70 degree curved suction was verified and then the malleable suction could no longer be tracked.Each instrument was connected to different ports.All other instruments were also connected and able to be tracked without any issues.Another malleable suction was connected and also could not be tracked.The surgeon opted to complete the procedure with the other navigated instruments and not the malleable suction.There was no impact to the patient.
 
Manufacturer Narrative
Patient information was not provided.Device manufacture date was unavailable as no lot number was provided.Evaluation was not possible as the device was discarded by the site.The account team confirmed there have not been any other issues with instrument tracking.The system was functioning normally with no need for additional service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALLEABLE SUCTION MEDIUM, STANDARD TIP
Type of Device
EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC.
826 coal creek circle
louisville CO 80027 971
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC.
826 coal creek circle
louisville CO 80027 971
Manufacturer Contact
amos jarrette
826 coal creek circle
louisville, CO 80027
7208902082
MDR Report Key4358982
MDR Text Key5238175
Report Number1723170-2014-01365
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133665
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 09/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735016
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/29/2014
Initial Date FDA Received12/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-