Catalog Number 20709
Device Problems Occlusion Within Device (1423); Failure to Sense (1559)
Patient Problem No Patient Involvement (2645)
Event Date 12/15/2014
Event Type  malfunction  
Event Description
The customer contact report that during preventative maintenance testing at the user facility, the device did not alarm when a distal occlusion was present.There were no reports of adverse pt effects and no reported delay in critical therapies while the device was in clinical use.No medical interventions were reported.No additional info was provided.
 
Manufacturer Narrative
At this time the customer will not be returning the device for evaluation.The device will be repair at the user facility and returned to clinical service.If the device is received, a follow-up report will be submitted.This report represents all the info known by the reporter upon query by hospira personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LC PCA REFURB 802.11
Type of Device
80MEA
Manufacturer (Section D)
HOSPIRA COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer (Section G)
HOSPIRA COSTA RICA LTD.
zona franca global
la aurora heredia
CS  
Manufacturer Contact
juergen schmider, md, vp
275 n. field drive, dept. no. 097u
bldg. no. h2-1 east
lake forest, IL 60045
2242125740
MDR Report Key4360572
MDR Text Key5234466
Report Number9615050-2014-06546
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042800
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number20709
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/15/2014
Initial Date FDA Received12/19/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-