• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES PROPLEGE CORONARY SINUS CATHETER; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES PROPLEGE CORONARY SINUS CATHETER; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number PR9
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Sinus Perforation (2277)
Event Date 11/17/2014
Event Type  Injury  
Event Description
Edwards received information that a coronary sinus injury occurred during the use of a proplege coronary sinus catheter.
 
Manufacturer Narrative
The device was not retained by the hospital for evaluation.At this time, there is no allegation of product malfunction but only an adverse event associated with the use of an edwards device.Injuries to the coronary sinus are listed as a potential complication in the product ifu.The ifu further notes: warning: if resistance is met, stop and re-evaluate proplege device position.The proplege device should not be advanced if resistance is felt, as doing, so would cause the proplege device to bend or buckle.Aggressive advancement in an attempt to engage the ostium may result in perforation or other injury.Warning: aggressive advancement of the guidewire in an attempt to engage the ostium may result in perforation or other injury.No device malfunction is indicated in the report and the events reported are included in the labeling.No actions are needed at this time this type of event is a known potential complication and/or adverse event, is included in the labeling/training materials, and in the information we provide to help avoid the issue.All labeling and training appropriately address the risk.Manufacturing records could not be reviewed as a lot number is unknown.Trends will continue to be monitored through the use of edwards quality systems.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPLEGE CORONARY SINUS CATHETER
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
EDWARDS LIFESCIENCES
12050 lone peak parkway
draper UT 84020
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
12050 lone peak parkway
draper UT 84020
Manufacturer Contact
walt wiegand
12050 lone peak parkway
draper, UT 84020
8015655200
MDR Report Key4361903
MDR Text Key20062996
Report Number3008500478-2014-00152
Device Sequence Number1
Product Code DWF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPR9
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/21/2014
Initial Date FDA Received12/24/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-