It was reported that on (b)(6) 2014, medics responded to a (b)(6) year old male patient in cardio pulmonary arrest.After the autopulse platform was removed from the carry case and the patient was placed on the device, it was noticed that the housing at the head had completely separated from the platform.They attempted to use the platform but it did not power on.The crew reverted to manual cpr (exact length of time was not provided).No adverse patient sequelae was reported.No further information was provided.
|
Investigation results for the returned platform as follows: visual inspection was performed and the reported complaint of the platform covers being damaged was confirmed.The following was observed: the top cover, motor cover, encoder cover, patient restraint assembly, restraint pin assembly and battery cover were damaged.From the condition of the returned unit, the damages appear to have been due to wear and tear.The reported complaint of the platform not powering on to perform compressions could not be confirmed.The platform ran for 25 minutes with a test mannequin and an additional 30 minutes with a large resuscitation test fixture (lrtf), with no issues encountered.The platform passed functional testing.There were no user advisories observed on the reported event date of (b)(6) 2014.Based on the investigation, the part(s) identified for replacement were the top cover, motor cover, encoder cover, patient restraint assembly, restraint pin assembly and battery cover.In summary, the reported complaint of the platform covers being damaged was confirmed during visual inspection and attributed to wear and tear.The reported complaint of the platform not powering on to perform compressions could not be duplicated.The platform passed all functional testing.Following service, including replacement of the damaged parts, the platform passed all testing criteria.
|