• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY VISIAN TORIC ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY VISIAN TORIC ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS Back to Search Results
Model Number VTICMO12.1
Device Problems Inadequacy of Device Shape and/or Size (1583); Appropriate Term/Code Not Available (3191)
Patient Problem No Information (3190)
Event Date 10/20/2014
Event Type  Injury  
Event Description
The reporter indicated the surgeon implanted a 12.1mm vticmo12.1 implantable collamer lens in the patient's left eye (os) on (b)(6) 2014.The lens had an excessive vault and remains implanted.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch report will be submitted.
 
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4) - size incorrect for patient; (vaulting); (implanted).Device evaluated by manufacturer? no.Lens remains implanted.Evaluation method: work order search.Results: a lens work order search was performed and no similar complaints were found within the work order.Conclusions - (no conclusion can be drawn): based on the complaint history and work order search, a specific root cause of the event could not be determined.(b)(4).
 
Manufacturer Narrative
Medical review - review of the file indicates that the lens was not implanted in accordance with the dfu requirements (e.G.Patient age below 21 or over 45 years, acd below 2.8mm for icl/ticl and below 3.0mm for vicl/vticl, keratoconus, pregnant or nursing patients, patients with low/abnormal corneal endothelial cell density, fuchs dystrophy or other corneal pathology, patients who are amblyopic or blind in the fellow eye).Off-label use of the device, no clinical data can support the complaint event(s) or the effect(s) on the efficacy and safety of the device.This information has been communicated to the medical advisor of staar surgical and to the surgeon in case of severe deterioration of patient health.Based on the complaint history, work order search and medical review, a specific root cause of the event could not be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIAN TORIC ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
althea watson
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key4362771
MDR Text Key20779474
Report Number2023826-2014-01083
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date08/31/2017
Device Model NumberVTICMO12.1
Other Device ID NumberDIOPTER UNK
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/18/2015
Initial Date FDA Received12/26/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/17/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/24/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE MODEL AND LOT NUMBER UNK; INJECTOR MODEL AND LOT NUMBER UNK
Patient Age27 YR
-
-