Model Number M00565100
Device Problems Material Deformation (2976); Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/05/2014
Event Type  Injury  
Event Description
It was reported to boston scientific corporation that a wallflex enteral colonic stent was implanted in the rectum during a colon stent deployment procedure performed on (b)(6) 2014.According to the complainant, the stent was implanted to treat a malignant stricture as a bridge to surgery (bts).Reportedly, the patient's anatomy was tortuous.During the procedure, the physician deployed the stent, but noted on endoscopic image that a stent wire was protruding through the delivery system.The physician pulled the delivery system and the stent was released.The proximal wires tore and the physician then attached the deformed wires to the proximal end of the stent using clips.The physician noted that the stent was fully deployed and expanded but was not deployed at its intended location.The stent remains implanted.The physician reported there was no damage noted to the delivery system.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
 
Manufacturer Narrative
Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.Reported event of stent placement / positioning problem.Reported event of stent broken.The complainant indicated that the device remains implanted and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX? COLONIC
Type of Device
STENT, COLONIC, METALIC, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key4363748
MDR Text Key5285015
Report Number3005099803-2014-04040
Device Sequence Number1
Product Code MQR
Combination Product (y/n)N
PMA/PMN Number
K061877
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/05/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/16/2016
Device Model NumberM00565100
Device Catalogue Number6510
Device Lot Number17053312
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2014
Initial Date FDA Received12/29/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/19/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-