MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG BEQ-HMOD70000-USA#QUADROX-ID AD.O.FIL.; DIFFUSIVE MEMBRANE OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG BEQ-HMOD70000-USA#QUADROX-ID AD.O.FIL.; DIFFUSIVE MEMBRANE OXYGENATOR Back to Search Results
Device Problems Fluid/Blood Leak (1250); Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Event Description
(b)(4).
 
Manufacturer Narrative
Maquet cardiopulmonary is aware of similar complaints.The devices displayed a similar malfunction which were tested and evaluated under an optical microscope.Delamination of some gas fibers was observed which allowed for the priming solution or blood to flow inside the gap between the gas fibers and polyurethane.Gravity then allowed for passage to the gas exiting path along the housing.The most probable root-cause is the delamination of the hollow gas fibers from the polyurethane potting area.A review of the quality control process confirms that 100% functional inspection for leakage was performed during production.Maquet cardiopulmonary (b)(4) has initiated an internal process (b)(4) to address the appropriate corrective and preventive action.A supplemental medwatch will be submitted when new info becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BEQ-HMOD70000-USA#QUADROX-ID AD.O.FIL.
Type of Device
DIFFUSIVE MEMBRANE OXYGENATOR
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer Contact
michael campbell
kehler strasse 31
rastatt 76437
GM   76437
2229321132
MDR Report Key4365441
MDR Text Key21329935
Report Number8010762-2014-01348
Device Sequence Number1
Product Code DTM
Combination Product (y/n)N
PMA/PMN Number
K101153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Remedial Action Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Other
Initial Date Manufacturer Received 11/24/2014
Initial Date FDA Received12/16/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2014
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-