Model Number 1884068HS |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/03/2014 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned on (b)(4) 2015.Analysis was completed on (b)(4) 2015.The product analysis found that for all three devices (2 samples of 1884560hs and 1 sample of 1884068hs), there was evidence of biologicals from visual inspection and reactivity with hydrogen peroxide.The hub and a portion of the inner shaft were separate from the rest of the devices.The inner shaft was broken 0.62" from the distal face of the inner hub which would have resulted in the reported malfunction.The break point corresponds to the proximal end of the outer tube in the front hub.When viewed under magnification, there were striations at the break point indicating metal on metal contact and an uneven break.The information indicates excess pressure was applied during use which caused deformation of the locking area, which then caused the inner shaft and outer tube to rub together until the inner shaft broke.
|
|
Event Description
|
It was reported that during surgery, 3 high-speed burs broke at the bottom of the shaft where the blade connects to the hub.There was no harm to the patient.The reporter was unable to confirm if the break resulted in any fragments in the surgical field.
|
|
Manufacturer Narrative
|
This device is used for therapeutic purposes.Concomitant device: bur 1884560hs; 510k: exempt; lot# 020680560 ; manufactured date: 2013-03-23 ; use-by date: 2021-03-21.(b)(4).The product has not been returned for analysis.Method: no testing methods performed.(b)(4).
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|