Model Number 3799 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Result: the report from the field was confirmed.Visual inspection found two contacts from the terminal end of a lead stuck in the bottom port of the header.The stuck terminal contacts were removed, the ipg was then charged to full and it was tested to manufacturing specifications using the autotester; all tests passed.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
The reported event occurred at the end of the patient's (b)(6) trial procedure.While removing the lead from the ipg header, one of the contacts (electrodes) came apart and remained inside the ipg header block.Per the physician, the set screw was backed out all the way.The event date is unknown at this time.
|
|
Manufacturer Narrative
|
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|