ST. JUDE MEDICAL - NEUROMODULATION PRODIGY IPG; SCS IPG Back to Search Results
Model Number 3799
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Result: the report from the field was confirmed.Visual inspection found two contacts from the terminal end of a lead stuck in the bottom port of the header.The stuck terminal contacts were removed, the ipg was then charged to full and it was tested to manufacturing specifications using the autotester; all tests passed.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
The reported event occurred at the end of the patient's (b)(6) trial procedure.While removing the lead from the ipg header, one of the contacts (electrodes) came apart and remained inside the ipg header block.Per the physician, the set screw was backed out all the way.The event date is unknown at this time.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODIGY IPG
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano, tx
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4366636
MDR Text Key20461148
Report Number1627487-2014-21789
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/01/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2015
Device Model Number3799
Device Lot Number4260265
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/01/2015
Initial Date FDA Received12/29/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/13/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-