ST. JUDE MEDICAL, INC. BRK-1? XS TRANSSEPTAL NEEDLE, 71 CM LENGTH Back to Search Results
Model Number G407209
Device Problem Unexpected Therapeutic Results (1631)
Patient Problem Pericardial Effusion (3271)
Event Date 12/05/2014
Event Type  Injury  
Event Description
During an electrophysiology procedure using a brk transseptal needle, a pericardial effusion occurred.Transseptal puncture with a brk transseptal needle was attempted multiple times without success.It was difficult to visualize the fossa ovalis due to a dilated aorta.An echocardiogram revealed a pericardial effusion.The patient remained asymptomatic for approximately 40 minutes and the procedure was aborted.The pericardial effusion resolved spontaneously with no intervention.There was no performance issue with the brk transseptal needle.
 
Manufacturer Narrative
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.Our investigation was limited to the review of the device history record, which showed that each manufacturing and inspection operation was performed and indicated complete in accordance with sjm specifications and procedures.Based on the information received, the cause of the reported incident could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK-1? XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TRANSSEPTAL NEEDLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
14901 deveau place
minnetonka MN 55345
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
14901 deveau place
minnetonka MN 55345
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key4367002
MDR Text Key5260347
Report Number3005188751-2014-00152
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K072278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/05/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2017
Device Model NumberG407209
Device Catalogue NumberG407209
Device Lot Number4659031
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2014
Initial Date FDA Received12/29/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SWARTZ BRAIDED TRANSSEPTAL INTRODUCER
Patient Outcome(s) Other;
Patient Age54 YR
Patient Weight93
-
-