BAXTER HEALTHCARE CORPORATION ABACUS®; ABACUS® TPN CALCULATION SOFTWARE, PRODUCT CODE: JQP Back to Search Results
Model Number 8300-0169
Device Problems Computer Software Problem (1112); Improper or Incorrect Procedure or Method (2017); Programming Issue (3014)
Patient Problem No Patient Involvement (2645)
Event Date 12/01/2014
Event Type  No Answer Provided  
Event Description
On (b)(6) 2014, baxter (b)(6) was informed of an incident occurring at (b)(6) healthcare of (b)(6) involving our abacus® software.Abacus® is our windows®-based order entry software used with our em2400 exactamix compounder.The exactamix compounder is an automated pumping system that compounds multiple sterile ingredients into a finished solution, and is designed for the preparation of total parenteral nutrition (tpn).While entering a test order the customer used "repeat selected" on the first order after modifying the template.The tpn contained double the requested volumes of multi-trace, m.V.I.And calcium gluconate.The doubling of the ingredients activated multiple oem and institutional warning limits.Baxter technical support instructed the customer on how to correct their set-up in order to eliminate the problem.There was no adverse event involved and no intervention was required; however, we are filing a mdr report because, if repeated, over delivery of an ingredient may require intervention or contribute to a serious patient injury or possible death.
 
Manufacturer Narrative
Evaluation: method: computer software performance tests conducted.· the software engineering team conducted extensive testing in order to duplicate and correct the problem.Results: software problem.· this issue is being investigated by baxter software engineering through a corrective action preventive action (capa).Conclusions: software / firmware caused event.· this issue is being investigated by baxter software engineering through a corrective action preventive action (capa).Should additional information regarding this event become available, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABACUS®
Type of Device
ABACUS® TPN CALCULATION SOFTWARE, PRODUCT CODE: JQP
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
9540 s. maroon circle
ste 400
englewood CO 80112
Manufacturer (Section G)
BAXTER HEALTHCARE CORPORATION
14445 grasslands drive
englewood CO 80112
Manufacturer Contact
lou fearing
9540 s. maroon circle
ste 400
englewood, CO 80112
MDR Report Key4367005
MDR Text Key5290513
Report Number1419106-2014-00017
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 12/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Pharmacist
Device Model Number8300-0169
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/29/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-