Catalog Number BLACKMAX-N_LL
Device Problem Air Leak (1008)
Patient Problem No Patient Involvement (2645)
Event Date 12/11/2014
Event Type  malfunction  
Event Description
It was reported that during pre-surgery, it was observed that the motor device had an air leak.There were no delays to the planned surgical procedure as an identical spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Additional narrative: the actual device was returned and is currently pending evaluation.Once reliability engineering evaluates the device, a follow-up medwatch will be submitted.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the outer hose was disconnected from the pressure release valve.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to excessive force on the hose, which was misuse and/or abuse.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLACKMAX-NEURO
Type of Device
MOTOR, DRILL, PNEUMATIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key4367267
MDR Text Key5116826
Report Number1045834-2014-15660
Device Sequence Number1
Product Code HBB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK831756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberBLACKMAX-N_LL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/21/2015
Initial Date FDA Received12/29/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/13/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-