Synthes is submitting this report as a result of remediation activities.Device listed in this report is used for treatment, not diagnosis.Any additional information received regarding this event after filing this report shall be filed on a supplemental mdr.A review of the device history shows revealed no complaint related anomalies.Subject device was received with a lag screw.An inspection of the device shows damage to the slotted end of the shaft.The slotted end of the shaft is slightly mushroomed and appears to be out of round.The slot also shows deformation due to the damage.No other damage is noted on the part.The damage to the lag screw occurred post manufacturing, as all pertinent features to this complaint are conforming.Therefore the complaint against the lag screw is indeterminate.
|