NIHON KOHDEN CORPORATION ZM-920PA; TELEMETRY TRANSMITTER Back to Search Results
Device Problems Overheating of Device (1437); Device Stops Intermittently (1599); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/21/2014
Event Type  malfunction  
Event Description
Customer stated that the zm-920pa telemetry transmitter had intermittent signal loss.The device also gets hot while powered on.Not on a pt.Imp ref #(b)(4).
 
Manufacturer Narrative
Although the device was not on a pt, if the malfunction were to occur there is likelihood that a pt could suffer a skin burn.The intermittent signal loss is easily detected by the central monitoring system (cns) and if the malfunction should recur, it is not likely to cause or contribute to death or injury.Device repaired and returned to customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-920PA
Type of Device
TELEMETRY TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
Manufacturer Contact
1-31-4 nishiochia, shinjuku-ku
tokyo 161-8-560
7050350
MDR Report Key4370530
MDR Text Key5236830
Report Number8030229-2014-00089
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
K945578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Type of Report Initial
Report Date 11/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Initial Date Manufacturer Received 11/28/2014
Initial Date FDA Received12/17/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2003
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-