STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 10° X3 INSERT 36MM ID; IMPLANT Back to Search Results
Catalog Number 623-10-36G
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Code Available (3191)
Event Date 12/12/2014
Event Type  Injury  
Event Description
Revision due to dislocation on left hip.Liner and head removed.
 
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Additional information was requested and if it becomes available will be submitted in a supplemental report.Not returned.
 
Manufacturer Narrative
The patient is (b)(6) in height.An event regarding dislocation involving a trident liner was reported.The event was not confirmed.Method and results: device evaluation and results: a visual, functional and dimensional inspection was not performed as the device was not returned.Medical records received and evaluation: insufficient information was received for review with a clinical consultant.Device history review: all devices in the reported lot were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of device, operative reports, xrays, patient history & follow-up notes are needed to investigate this event further.
 
Event Description
Revision due to dislocation on left hip.Liner and head removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 10° X3 INSERT 36MM ID
Type of Device
IMPLANT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
beverly lima
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key4372363
MDR Text Key5291738
Report Number0002249697-2014-05050
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K033716
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2019
Device Catalogue Number623-10-36G
Device Lot NumberMNL70E
Other Device ID NumberSTERILE LOT# MSGNL23H5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/13/2015
Initial Date FDA Received12/30/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/11/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/24/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age59 YR
Patient Weight91
-
-