SPECTRANETICS CORPORTATION SPECTRANETICS LEAD LOCKING DEVICE; LLD EZ Back to Search Results
Model Number 518-062
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Low Blood Pressure/ Hypotension (1914)
Event Date 12/15/2014
Event Type  Injury  
Event Description
Lead management case to extract one mdt 5076 ra lead (impl.10 months) due to noise.The physician inserted an lld ez and began the procedure with a 14f glidelight.Before the physician made the turn to the ra/svc junction the lead stretched out and came off the ra appendage.Approximately 7-8 minutes after removal the patient's blood pressure began to drop.A pericardiocentesis was performed and the patient stabilized within 10 minutes without further intervention.The physician believed that the effusion was caused by a small tear in the ra appendage caused from the ra lead screw.This is being reported because the lld was used as the traction rail inside of the cardiac lead.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD EZ
Manufacturer (Section D)
SPECTRANETICS CORPORTATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS CORPORTATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
sondra chandler rn
9965 federal drive
colorado springs, CO 80921
7194472530
MDR Report Key4372441
MDR Text Key18896225
Report Number1721279-2014-00221
Device Sequence Number1
Product Code DRB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043401
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date09/03/2016
Device Model Number518-062
Device Catalogue Number518-062
Device Lot NumberFLP14J03A
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/15/2014
Initial Date FDA Received12/30/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/03/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM; MDT 5076 CARDIAC LEAD (IMPL. 10MO); SPECTRANETICS 14F GLIDELIGHT LASER SHEATH
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age44 YR
-
-