Model Number 0800
Device Problems Ambient Temperature Problem (2878); Device Operates Differently Than Expected (2913); Moisture or Humidity Problem (2986)
Patient Problems Incontinence (1928); Necrosis (1971)
Event Date 11/25/2014
Event Type  Injury  
Event Description
Hill-rom received a report from the account alleging the pt has a pre-existing stage 2 sacral wound that progressed to a stage 4.The wound care nurse stated an incision was made to remove the necrotic tissue and the wound has been receiving daily antibiotic ointment treatments along with cleaning with saline solution and the use of aqua seal silver dressing.The bed was located in the pt's home.This report was filed in our complaint handling system as complaint # (b)(4).The wear of microspheres to the point that they will not fluidize properly is multifaceted and their replacement is a practical occurrence that is inherent with use of the bed.Several factors come into account, environment, use and care of the pt.Hill-rom recommends not exceeding an ambient room temperature of 75 degrees f.The microspheres can handle limited amounts of fluids passing through the filter sheet.Excessive humidity, incontinence and bodily fluids will saturate the beads and hamper fluidization.Treatment from caregivers such as petroleum based topical ointments and silver compounds will ruin the coating on the beads and permanently destroy their fluidizing properties.
 
Manufacturer Narrative
The hill-rom technician found the micro spheres were over saturated with moisture and needed to be replaced.The technician replaced the microspheres to resolve the issue.The reported injury is serious in nature per fda definition.Moisture in beads was not related to a malfunction of the device.Development of pressure ulcers is multifactorial and cannot be only attributed to performance of the surface.Risk factors include protein-calorie malnutrition, microclimate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy.Position changes are key to pressure sore prevention and treatment.These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITRON AT HOME
Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key4372881
MDR Text Key5263186
Report Number1824206-2014-02882
Device Sequence Number1
Product Code INX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K942184
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model Number0800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/25/2014
Initial Date FDA Received12/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age60 YR
Patient Weight54
-
-