Customer reported by fax: "buffy coat exceeded at 1637ml wbp.Hematocrit froze at 12% on the screen.Cut bc early, then noticed clear fluid coming from hct sensor and a pool of water (small amount).No more noticed, could not find where this had come from.Treatment completed.Patient was feeling fine".Treatment was in single needle mode, collect rate 25ml/min.Clinical services specialist called customer on (b)(4) for more information: customer noticed some fluid ("few drops") in the hematocrit sensor when she lifted the cuvette out of the sensor housing.She tried to find a cause for the hct curve to freeze at 12%, but could not see anything strange.Customer confirmed it was clear fluid, as a water drop, certainly not blood.Customer carefully examined lines and pto for leaks, but there were none.Customer continued the procedure and during photoactivation placed a tissue under the lines in the area of the fluid.All remained dry, the treatment was therefore completed, the blood returned to the patient.The customer will return the smart card for investigation.There was no service order generated.
|