• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA INC. PUMP IN STYLE ADVANCE BREAST PUMP GO TOTE; HGX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA INC. PUMP IN STYLE ADVANCE BREAST PUMP GO TOTE; HGX Back to Search Results
Model Number 57063/9207010
Device Problems Crack (1135); Protective Measures Problem (3015)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/15/2014
Event Type  malfunction  
Event Description
The customer originally reported to customer service that the transformer housing for her pns was cracked, but not coming apart.Upon receipt at medela the transformer was inspected and found to be breached, which is a safety risk.
 
Manufacturer Narrative
A replacement power supply was sent to the customer.Upon receipt of the product an evaluation showed that the housing was breached, exposing internal electronics, which is a safety risk.Per qe evaluation, the transformer housing appears to have been broken and repaired by the end user.This issue with the damaged transformer housing for the pump in style device was addressed in investigation (b)(4).The investigation found that the transformers are being damaged during shipment from the manufacturer to medela.This damage is causing the plastic housing to fail prematurely when subjected to normal use and foreseeable misuse.The packaging used by the manufacturer to ship the transformers to medela is not robust enough to handle all of the potential shipping, handling, and abuse conditions that could arise from logistics of the consolidation process.As a result of the investigation, the shipping and consolidation process has been modified to reduce the handling and potential for double stacking of the skids.The shipping packaging strength has also been increased to further protect the transformers during shipping.Evaluation summary: a visual examination of the power supply revealed the transformer housing was breached, exposing inner electrical circuitry.A visual examination of the cord revealed nothing unusual.The voltage across the dc plus was not available.The transformer was unsafe to plug in.The resistance across the ac blades measured as 54.4 ohms, which is normal and indicates that the thermal fuse did not blow.Resistance across the dc plug was measured open, which indicates that there is not a short in the dc cord.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE ADVANCE BREAST PUMP GO TOTE
Type of Device
HGX
Manufacturer (Section D)
MEDELA INC.
mchenry IL 60050
Manufacturer Contact
don alexander
1101 corporate dr
mchenry, IL 60050
8004358316
MDR Report Key4374485
MDR Text Key17946548
Report Number1419937-2014-00936
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 12/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number57063/9207010
Device Catalogue Number57063/9207010
Device Lot NumberREV L-0912
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/04/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/15/2014
Initial Date FDA Received12/24/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-