MEDTRONIC NEUROMODULATION INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
|
Back to Search Results |
|
Model Number 3625 |
Device Problems
Failure to Deliver Energy (1211); Migration or Expulsion of Device (1395)
|
Patient Problems
Therapeutic Effects, Unexpected (2099); Therapeutic Response, Decreased (2271)
|
Event Date 12/09/2014 |
Event Type
malfunction
|
Event Description
|
It was reported the trial patient had a loss of therapeutic effect.They felt stimulation faintly.Increasing to 10.0 v did not resolve the issue.The patient would feel stimulation and it would get less and less until they could not feel it.The issue occurred with both sides.The patient was getting up three times a night.It was later clarified the patient only had one lead.Sensation was lost due to the lead falling out.
|
|
Manufacturer Narrative
|
Concomitant products: product id neu_unknown_lead, lot # unknown, product type lead; product id neu_unknown_lead, lot # unknown, product type lead.(b)(4).
|
|