Data received 08/27/2015 by manufacturer with successive device evaluation 09/18/2015: lot number 6267.Serial number (b)(4).Approximately june 30, 2015, remaining linx device was explanted.Device was explanted simultaneous to patient's fundoplication being undone due to herniation of the fundoplication.Device was found in the correct position and geometry.A second fundoplication was completed after undoing the initial fundoplication and removal of the remaining linx device.Patient is reported as well.09/18/2015: device analysis conducted by torax medical engineering after product receipt.Gross analysis revealed no device anomalies atypical from an explanted device.Microscopic analysis revealed all device components and assembly exhibited normal characteristics of an explanted device.Force testing, length testing, and geometric analysis exhibited no anomalies.No conclusion relevant to experience determined.
|