ALERE SAN DIEGO, INC. CONSULT DIAGNOSTICS HCG CASSETTE; HCG PREGNANCY TEST Back to Search Results
Model Number FHC-A102-OBC554
Device Problems False Negative Result (1225); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2014
Event Type  malfunction  
Event Description
Caller alleging receiving false negative urine hcg results using hcg cassette for two patients.Hcg product received a negative result, patient went to emergency room and was evaluated as positive.Patient 2 (b)(6) - email from distributor alleges patient went into doctor's office with abdominal pain.The pregnancy test gave a false negative result and patient was sent to the emergency room where she was confirmed positive.Patient was 9 weeks pregnant.All of this took place on the same day and the patient and baby are in good condition now.No additional information was provided.Mdr #2027969-2014-01099 submitted to address the complaint for patient 1.
 
Manufacturer Narrative
Investigation conclusion: customer's observation was not replicated in-house with retention devices.Retention devices were tested with 20 miu/ml cutoff hcg urine control and 3 high level hcg urine controls, all results were positive at read time.No false negative were obtained.Manufacturing batch record review did not uncover any abnormalities.Root cause could not be determined without patient specimen in-house analysis.Based on the information available, there is no indication of a product deficiency.No corrective action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSULT DIAGNOSTICS HCG CASSETTE
Type of Device
HCG PREGNANCY TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
san diego CA
Manufacturer Contact
ya-ling king
9975 summers ridge rd
san diego, CA 92121
8588052084
MDR Report Key4384791
MDR Text Key5367377
Report Number2027969-2014-01100
Device Sequence Number1
Product Code JHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFHC-A102-OBC554
Device Lot NumberHCG4010136
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2014
Initial Date FDA Received12/31/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-