Event Description
The customer reported that their device appeared locked up by showing a partial solid white screen with vertical lines across the screen.The device likely could not function in this state.There was no patient use associated with the reported issue.
 
Manufacturer Narrative
(b)(4).Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Manufacturer Narrative
Physio-control evaluated the customer's device and verified the reported issue.Physio then replaced both the system controller (sc) and user interface (ui) pcb assemblies and completed other, unrelated, repairs.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.Physio further examined the removed ui pcb assembly and determined that the cause of the reported issue was an electrically shorted integrated circuit (ic) chip, designator u8.The sc pcb assembly was an ancillary part and there was no failure of the assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4391347
MDR Text Key5124280
Report Number3015876-2015-00023
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2015
Is the Reporter a Health Professional? No
Device Age13 YR
Event Location Hospital
Initial Date Manufacturer Received 03/25/2015
Initial Date FDA Received01/07/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/03/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/24/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-