Catalog Number ACE36E
Device Problems Loss of Power (1475); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/30/2014
Event Type  malfunction  
Event Description
It was reported that during an unknown procedure, after two-three activations of a new ace probe (opened in this surgery), the generator started showing a replace instrument error and did not work even after the troubleshooting steps were followed.Unknown how case was completed.There were no adverse consequences for the patient.
 
Manufacturer Narrative
(b)(4).Information asked for but unknown or not provided during initial contact.Should the information be provided later, a supplemental medwatch will be sent.The device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched.This blade tip portion may have broken off of the device during transport to our analysis site.The reported complaint was for the generator started showing a replace instrument error.The device was functionally tested with a generator.During functional testing on gen11 an alert screen was displayed.A probable cause of an alert screen is blade damage.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRACISION HARMONIC ACE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 0096 9
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 0096 9
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4397027
MDR Text Key18730723
Report Number3005075853-2015-00220
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K042777
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/31/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/27/2017
Device Catalogue NumberACE36E
Device Lot NumberJ90F5G
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/07/2015
Initial Date FDA Received01/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment