• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MALTA ACCESS; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MALTA ACCESS; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number MMC2071B
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 12/16/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned for evaluation.A visual inspection and an underwater leak test were performed.During the leak testing it was identified that there was a leak between the chamber bushing and the tube.Upon closer inspection it was identified that this was due to a lack of solvent between two components.The lack of solvent is the result of a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an access administration set was leaking "just below the 2 drip chambers".There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - MALTA
a47 industrial estate, malta b
marsa
MT 
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate, malta b
marsa
MT  
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4398165
MDR Text Key12892200
Report Number1416980-2015-00935
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/17/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2019
Device Catalogue NumberMMC2071B
Device Lot Number14G31V141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/21/2015
Initial Date FDA Received01/08/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/09/2015
02/19/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-