Catalog Number EMAX2PLUS |
Device Problems
Material Protrusion/Extrusion (2979); Noise, Audible (3273)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/18/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during an unspecified surgical procedure, it was observed that the motor device started to pop out of the handpiece and then started making loud noise and smoke.There were no delays in the surgical procedure as a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device would not lock the cutter device.It was determined that the cutter would not lock because the locking mechanism was stripped (powerbroken).Therefore, the pre-test could not be completed.The reported conditions were confirmed.It was determined that the cause of the powerbroken was failure to follow directions for use and running the device in the safe or load position.The assignable root cause was determined to be due to misuse, abuse and possibly user error.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|