Catalog Number 0986000000 |
Device Problem
Sparking (2595)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/01/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that the cast vacuum motor is sparking during a cast removal procedure at user facility.The procedure was completed successfully.No delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
The reported event, it was throwing sparks, was duplicated.Through inspection, the service technician noted that the castvac was sparking and making noise due to a motor failure.Upon disassembly, the technician noted that the castvac was full of dust suffocating the motor.The motor was replaced along with other preventative maintenance returned to the customer after passing the final inspection.
|
|
Event Description
|
It was reported that the cast vacuum motor is sparking during a cast removal procedure at user facility.The procedure was completed successfully.No delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
The device has not been received at the manufacturer for testing. an evaluation will be conducted upon receipt of the device, and a follow-up report will be submitted after the quality investigation is complete.
|
|