DEPUY SYNTHES POWER TOOLS ELECTRIC SYSTEMS FOOT CONTROL; MOTOR, DRILL, ELECTRIC - FOOT CONTROL
|
Back to Search Results |
|
Catalog Number E-FP_LL |
Device Problems
Device Displays Incorrect Message (2591); Material Protrusion/Extrusion (2979)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/19/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during an unspecified surgical procedure on the knee, it was observed that the foot control device had an e9 error code.It was further reported that wires were exposed on the back of the device.There were no delays to the surgical procedure as an identical spare device was available for use.The surgery was completed successfully with the spare device.There was patient involvement.There were reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the cord was damaged at the back of the device.Therefore, the reported condition was confirmed.It was determined that this was indicative of an extreme pulling motion which caused the cord to be ripped out and/or from pulling the device around by the cord.The assignable root cause was determined to be due to misuse, abuse, and/or user error.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|