MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS ELECTRIC SYSTEMS FOOT CONTROL; MOTOR, DRILL, ELECTRIC - FOOT CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS ELECTRIC SYSTEMS FOOT CONTROL; MOTOR, DRILL, ELECTRIC - FOOT CONTROL Back to Search Results
Catalog Number E-FP_LL
Device Problems Device Displays Incorrect Message (2591); Material Protrusion/Extrusion (2979)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2014
Event Type  malfunction  
Event Description
It was reported that during an unspecified surgical procedure on the knee, it was observed that the foot control device had an e9 error code.It was further reported that wires were exposed on the back of the device.There were no delays to the surgical procedure as an identical spare device was available for use.The surgery was completed successfully with the spare device.There was patient involvement.There were reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the cord was damaged at the back of the device.Therefore, the reported condition was confirmed.It was determined that this was indicative of an extreme pulling motion which caused the cord to be ripped out and/or from pulling the device around by the cord.The assignable root cause was determined to be due to misuse, abuse, and/or user error.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRIC SYSTEMS FOOT CONTROL
Type of Device
MOTOR, DRILL, ELECTRIC - FOOT CONTROL
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key4414550
MDR Text Key5377158
Report Number1045834-2015-10056
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberE-FP_LL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2014
Initial Date Manufacturer Received 12/22/2014
Initial Date FDA Received01/13/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-