• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE FLORIDA OPERATIONS PERFECTO2 V WITH SENSOR 9153650799; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE FLORIDA OPERATIONS PERFECTO2 V WITH SENSOR 9153650799; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number IRC5PO2
Device Problems Break (1069); Overheating of Device (1437); Loss of Power (1475); Shipping Damage or Problem (1570); Insufficient Flow or Under Infusion (2182); Out-Of-Box Failure (2311); Device Dislodged or Dislocated (2923); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Dealer states the air baffle was improperly installed from the factory and it completely dislodged in shipping, which made the unit run very hot, shutting off eventually but severely deforming air baffle, base and ruining sieve beds, which are causing low o2.No further information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFECTO2 V WITH SENSOR 9153650799
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44036
8003336900
MDR Report Key4435188
MDR Text Key5378145
Report Number1031452-2015-03483
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 01/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberIRC5PO2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/12/2015
Initial Date FDA Received01/21/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-