Additional device component: ami 9700 console ((b)(4)).Device eval summary: test scans were conducted on a 3.67 cm aortic phantom.The probe did not produce an image, and the measurement given was the default of less than 3 cm.The scan cable was evaluated on the cable eye test fixture, and an open circuit in the signal wire was found (the circf j1 pin 1 to circf j2 pin 3).This prevented the probe from sending an ultrasound image to the console to produce a measurement.After replacing the scan cable, the unit scanned correctly, measuring 3.7 cm on a 3.67 cm aortic test phantom.
|