Model Number 9-ASD-034
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Information (3190)
Event Type  Injury  
Event Description
A 34mm amplatzer septal occluder (aso) was implanted and it was then determined it was too small for the patient.The patient was referred for surgery where the aso was explanted.
 
Manufacturer Narrative
(b)(4).The 34mm aso was returned to sjm and decontaminated.The occluder was grossly and microscopically examined, and no anomalies were found.It met dimensional specifications when measured with a caliper.No additional testing was performed.The device history record for this product was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with sjm specifications and procedures.The results of this investigation confirmed the aso met all dimensional specifications when analyzed at sjm.There was no evidence to suggest there was an intrinsic defect in the occluder, and the cause for the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key4447716
MDR Text Key18662367
Report Number2135147-2014-00128
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeMO
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/30/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date03/06/2017
Device Model Number9-ASD-034
Device Catalogue Number9-ASD-034
Device Lot Number1203078867
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/30/2014
Initial Date FDA Received01/23/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age11 YR
Patient Weight35
-
-