Model Number T175
Device Problems Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/24/2014
Event Type  malfunction  
Event Description
Boston scientific received information that information that this implantable cardioverter defibrillator (icd) delivered anti-tachycardia pacing (atp) which accelerated the patient's rhythm into the ventricular tachycardia (vt) zone where shock therapy was delivered.Therapy was exhausted.It was unclear whether the patient's rhythm was supraventricular tachycardia (svt) or a true ventricular tachycardia (vt).Boston scientific technical services discussed potential programming changes.There were no adverse patient effects.The device remains in service.
 
Manufacturer Narrative
(b)(4).As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 55112
6515824786
MDR Report Key4452957
MDR Text Key21560966
Report Number2124215-2015-01150
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 10/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/01/2009
Device Model NumberT175
Other Device ID NumberVITALITY 2 VR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 10/24/2014
Initial Date FDA Received01/26/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/03/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0185; T175
Patient Age67 YR
-
-