Model Number H30120019680 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/16/2014 |
Event Type
malfunction
|
Event Description
|
Same case as mdr 2134265-2015-00348.It was reported that unintended rf energy was delivered to the patient.During an ablation procedure, 10hz noise was seen on the ablation channel when no rf energy was applied.Somehow there was pacing captured in the atrium for the ablation catheter (10hz), indicating the ablation catheter was sending energy into the patient.No patient complications were reported and the patient is in good condition.No additional treatment was required.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the results of the testing indicated that the returned carto cable as well as the associated clearsign amplifier passed incoming inspection tests, including noise and leakage testing.The cable shielding connection was verified by measuring the capacitance and impedance and results were similar to the control cable.The possibility of interaction effects between the two components as a cause was eliminated by testing the returned complaint amplifier and cable together and found no issues.Both devices performed as expected.The batch number is unknown and the manufacturing records for the complaint device could not be reviewed.The most probable cause of the reported difficulties may be due interaction with another device.(b)(4).
|
|
Event Description
|
Same case as mdr 2134265-2015-00348.It was reported that unintended rf energy was delivered to the patient.During an ablation procedure, 10hz noise was seen on the ablation channel when no rf energy was applied.Somehow there was pacing captured in the atrium for the ablation catheter (10hz), indicating the ablation catheter was sending energy into the patient.No patient complications were reported and the patient is in good condition.No additional treatment was required.
|
|
Search Alerts/Recalls
|