• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE CLEVELAND STREET SITZ BATH 1616700692; BATH, SITZ, NONPOWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE CLEVELAND STREET SITZ BATH 1616700692; BATH, SITZ, NONPOWERED Back to Search Results
Model Number IH3602GXL
Device Problems Leak/Splash (1354); Smoking (1585)
Patient Problem No Information (3190)
Event Date 01/26/2015
Event Type  malfunction  
Event Description
(b)(6) 2015- no injury alleged.Malfunction alleged.Per staff, around the door there is water leaking.Mdr filed.Additional reportable malfunction for this device; (b)(4) - hot (electrical) smell and smoke.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITZ BATH 1616700692
Type of Device
BATH, SITZ, NONPOWERED
Manufacturer (Section D)
INVACARE CLEVELAND STREET
899 cleveland street
elyria OH 44036
Manufacturer (Section G)
INVACARE CLEVELAND STREET
899 cleveland street
elyria OH 44036
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44036
8003336900
MDR Report Key4477356
MDR Text Key21328549
Report Number1219590-2015-00034
Device Sequence Number1
Product Code ILJ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 01/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberIH3602GXL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/27/2015
Initial Date FDA Received02/03/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-