MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
|
Back to Search Results |
|
Model Number 3058 |
Device Problems
Unable to Obtain Readings (1516); Device Operates Differently Than Expected (2913)
|
Patient Problems
Incontinence (1928); Therapeutic Effects, Unexpected (2099); Urinary Tract Infection (2120)
|
Event Type
Injury
|
Event Description
|
It was reported the patient was unable to adjust stimulation.Putting batteries in the programmer resolved that issue.The patient never had therapeutic effect.They could feel stimulation, but were urinating like prior to implant.The issue started a few months prior to call.The patient had to wear a diaper and pad.A week later, it was determined a urinary tract infection was the cause of the issue.Adjusting settings had not resolved the issue.A urine culture tested positive.Unipolar impedance measurements for electrodes zero and one showed ¿???¿.All other impedances were within normal limits.The patient had cancelled their follow-up appointment.No outcome was reported regarding this event.Further follow-up is being conducted to obtain this information.If additional information is received, a follow-up report will be sent.
|
|
Manufacturer Narrative
|
Concomitant products: product id 3093-28, lot# v765390, implanted: (b)(6) 2011, product type lead; product id 3037, serial# (b)(4), product type programmer, patient.(b)(4).
|
|