• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Unable to Obtain Readings (1516); Device Operates Differently Than Expected (2913)
Patient Problems Incontinence (1928); Therapeutic Effects, Unexpected (2099); Urinary Tract Infection (2120)
Event Type  Injury  
Event Description
It was reported the patient was unable to adjust stimulation.Putting batteries in the programmer resolved that issue.The patient never had therapeutic effect.They could feel stimulation, but were urinating like prior to implant.The issue started a few months prior to call.The patient had to wear a diaper and pad.A week later, it was determined a urinary tract infection was the cause of the issue.Adjusting settings had not resolved the issue.A urine culture tested positive.Unipolar impedance measurements for electrodes zero and one showed ¿???¿.All other impedances were within normal limits.The patient had cancelled their follow-up appointment.No outcome was reported regarding this event.Further follow-up is being conducted to obtain this information.If additional information is received, a follow-up report will be sent.
 
Manufacturer Narrative
Concomitant products: product id 3093-28, lot# v765390, implanted: (b)(6) 2011, product type lead; product id 3037, serial# (b)(4), product type programmer, patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4478784
MDR Text Key18070326
Report Number3004209178-2015-01612
Device Sequence Number1
Product Code EZW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 01/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2013
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/12/2015
Initial Date FDA Received02/03/2015
Date Device Manufactured09/09/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age00079 YR
-
-