• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM AC POWER SUPPLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM AC POWER SUPPLY Back to Search Results
Catalog Number 295050-001
Device Problems Break (1069); Device Operates Differently Than Expected (2913); Electrical Power Problem (2925); Power Problem (3010)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 01/12/2015
Event Type  malfunction  
Event Description
The freedom a/c power supply was not in patient use.The freedom a/c power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that the freedom a/c power supply had a broken plug and would not function properly.This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom a/c power supply was not in patient use.In addition, the freedom driver has a redundant power source of onboard batteries.
 
Manufacturer Narrative
The freedom a/c power supply will be returned to syncardia for evaluation.The results of the investigation will provided in a supplemental mdr.
 
Manufacturer Narrative
(b)(4).
 
Event Description
This freedom a/c power supply was not in patient use.The freedom a/c power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that the freedom a/c power supply had a broken plug and would not function properly.Freedom ac power supply was returned to syncardia for evaluation.Visual inspection of the freedom ac power supply revealed that the cable connector cover was installed backwards.This is the root cause of the customer-reported inability to connect the ac power supply to a freedom power adaptor.The ac power supply cable was disassembled, the connector key slot was aligned with the connector cover key tab and the correct mating configuration was obtained.After the ac power supply cable was reassembled, the ac power supply passed all testing requirements.This issue was identified to be isolated to one lot that was inspected at syncardia.In response to an unrelated issue regarding power supply connectors coming apart, the inspector performed a verification to ensure the connectors' interior threads were properly mated with adhesive.The verification required the inspector to remove the connector covers.Therefore, it is possible that the cable connector cover was installed incorrectly during reassembly at syncardia during incoming inspection.Incoming inspection requirements were improved to clearly identify the correct removal and replacement of the connector cover.This failure mode poses a low risk to a patient because the freedom ac power supply was not in use by a patient at the time of the customer-reported issue.In addition, it would not prevent a freedom driver from performing its life-sustaining functions.The freedom driver is equipped with redundant power sources, including multiple rechargeable freedom onboard batteries and a car charger.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM AC POWER SUPPLY
Type of Device
AC POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key4484366
MDR Text Key16172729
Report Number3003761017-2015-00025
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,health professional,use
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number295050-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2015
Initial Date FDA Received01/29/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/03/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-