Catalog Number B-VIOLET
Device Problem Overheating of Device (1437)
Patient Problem No Patient Involvement (2645)
Event Date 01/12/2015
Event Type  malfunction  
Event Description
It was reported that during service and repair pre- testing, it was observed that the attachment device had high temperature.The event was not related to surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The device was returned for service, however did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to worn out bearings from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8.2CM ATTACHMENT, BLACK MAX
Type of Device
BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key4489832
MDR Text Key5329038
Report Number1045834-2015-10253
Device Sequence Number1
Product Code GFF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK974025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 01/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB-VIOLET
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/12/2015
Initial Date FDA Received02/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/29/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-