Catalog Number 4207000000 |
Device Problem
Fracture (1260)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/12/2015 |
Event Type
malfunction
|
Event Description
|
It was reported during service at manufacturer facility that a part of the handpiece chuck is broken off.No associated procedure, no patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
The device has been received, failure analysis is in progress; additional information will be submitted once the quality investigation is completed.Failure analysis is in progress.
|
|
Event Description
|
It was reported during service at manufacturer facility that a part of the handpiece chuck is broken off.No associated procedure, no patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
The reported event, part of chuck and yoke assembly was broken off, was confirmed.The piece that was broken off was all internal and nothing fell out of the handpiece.The chuck and yoke assembly was replaced with other suggested components.Preventative maintenance was performed and the handpiece was returned to the customer.
|
|