Catalog Number 121740500 |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
|
Patient Problems
Foreign Body Reaction (1868); Pain (1994)
|
Event Date 09/03/2013 |
Event Type
Injury
|
Event Description
|
Litigation alleges that the patient suffers from pain, discomfort, inflammation, and large amounts of toxic cobalt chromium metal ion particles to be released into the blood, tissue, and bone.Update received: 1/20/2015.The sales rep has reported the revision surgery.Patient was revised to address pain.Complaint was updated on 1/26/2015.
|
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
|
|
Manufacturer Narrative
|
Additional narrative: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|
Manufacturer Narrative
|
Depuy still considers this investigation closed at this time.
|
|
Event Description
|
Update 6/24/15-pfs and medical records received.After review of the medical records for mdr reportability, the revision operative note indicated pain, metal ion levels below 7ppb, and corrosion between the head and stem.The stem is being added to the complaint.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: udi: (b)(4).Added: b5, h6.
|
|
Event Description
|
Ppf alleges metallosis and elevated metal ion.
|
|
Search Alerts/Recalls
|