TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; SARNS CENTRIFUGAL SYSTEM Back to Search Results
Model Number 816572
Device Problems No Display/Image (1183); Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/15/2015
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the centrifugal controller unit display failed.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was verified by the field representative (fsr).The fsr disassembled the centrifugal control unit.The sr found the ribbon cable that goes from the display board to the pd board was disconnected.The customer has the unit pole mounted but not tightened down.The unit is loose enough where they can swivel it around.The unit most likely got jarred which probably caused the issue.The latch down tabs on either side of the cable connection were loose.When reinstalling the cable, the fsr put silicone sealant on each side latch.The fsr reassembled the unit, the display came on with no further issues.The fsr tested the unit.The unit operated to manufacturer specifications and was returned to clinical use.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4494308
MDR Text Key5570262
Report Number1828100-2015-00065
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816572
Device Catalogue Number816572
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/16/2015
Initial Date FDA Received02/05/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse