It was reported that during a thoracic outlet syndrome procedure, the device worked normally in the beginning of the procedure.After about fifteen minutes, it stopped working.The instrument was tightened-but failed to function.Visually there was a fracture on the blade.Another like device was used to complete the procedure.There were no adverse consequences for the patient reported.
|
(b)(4).Should the information be provided later, a supplemental medwatch will be sent.The device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched.This blade tip portion may have broken off of the device during transport to our analysis site.The reported complaint was for: "activation issues and blade damage".The device was functionally tested with a generator.During functional testing on gen11 an alert screen was displayed.A probable cause of the device stop activating and display an alert screen is blade damage.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.
|