Model Number 2090
Device Problems Device Emits Odor (1425); Failure to Power Up (1476); Noise, Audible (3273)
Patient Problem No Patient Involvement (2645)
Event Date 12/15/2014
Event Type  malfunction  
Event Description
It was reported that when powered on the programmer made a loud "whining" noise then a "snapping" noise and then display went black.There was a faint smell of electrical odor.The programmer would not turn back on.The programmer was returned for service.There was no patient involvement.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: analysis confirmed that the unit would not power up and that the power supply had a faint odor.The power supply was replaced.It was further noted that the hard drive was noisy as was the system fan, both were replaced and the hard drive imaged.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4505172
MDR Text Key5442449
Report Number2182208-2015-00653
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/15/2014
Initial Date FDA Received02/10/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
229047 SOFTWARE ANALYZER
-
-