• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LIMITED INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LIMITED INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT266
Device Problem Material Separation (1562)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/13/2015
Event Type  malfunction  
Event Description
A hospital in (b)(6), reported via a fisher & paykel healthcare field representative that an rt266 infant dual-heated evaqua2 breathing circuit came apart at the wye and would not connect back together.
 
Manufacturer Narrative
(b)(4).The complaint rt266 breathing circuit was not returned to fisher & paykel healthcare in (b)(4) for evaluation.Our investigation is based on our knowledge of the product and the information provided by the hospital.The hospital reported that the breathing circuit passed the initial leak test and that the wye detached while the breathing circuit was being handled.The hospital reported that no damage was observed on the wye piece.Without the return of the complaint device we are unable to determine what may have caused the reported event.All rt266 infant dual-heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The customer has confirmed that the subject breathing circuit was in use before the event occurred, which indicates that the circuit became damaged during use.The user instructions that accompany the rt266 state the following: - "check all connections are tight before use." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "set appropriate ventilator alarms.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, auckland 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, auckland 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
MDR Report Key4508956
MDR Text Key16991554
Report Number9611451-2015-00072
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Respiratory Therapist
Type of Report Initial
Report Date 01/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT266
Device Catalogue NumberRT266
Device Lot Number141020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/13/2015
Initial Date FDA Received02/11/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/20/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-