Model Number 722006
Device Problem Intermittent Continuity (1121)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Philips received a complaint from a customer that they had intermittently problems with both the foot switches for fluoroscopy and cine.
 
Manufacturer Narrative
(b)(4).The site reported the footswitch was not working.The field service engineer's visual review of the system showed the footswitch was bent and not working properly.Philips conducted its investigation based on info received.Based on available info, the cause could be related to the use of anti-fatigue mats under the footswitches since the footswitch was designed to be used on a solid surface.After replacing the footswitch, the reported problem with the system was solved.Philips is working on adding a metal plate under the footswitch to prevent bending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD20
Type of Device
IZI
Manufacturer (Section D)
PHILIPS HEALTHCARE
veenpluis 4-6
best 5680 DA
NL  5680 DA
Manufacturer (Section G)
PHILIPS HEALTHCARE
veenpluis 4-6
best 5680 DA
NL   5680 DA
Manufacturer Contact
dominic siewko
3000 minuteman rd
andover, MA 01810
9786597936
MDR Report Key4510877
MDR Text Key15237047
Report Number3003768277-2014-00056
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031333
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number722006
Device Catalogue Number722006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/29/2014
Initial Date FDA Received05/28/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-