Model Number 100
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the blood parameter monitor (bpm) took a long time to power up.The device was not changed out, as they used the bpm for the case.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
This complaint is related to mdr #1828100-2015-00133.This has been an ongoing issue for the past three months.The reported complaint was confirmed.During inspection of the bpm, it was found to have a damaged battery and battery acid had leaked into the unit effecting the unit to power on.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO CDI 100 MONITOR
Type of Device
CDI 100
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4511466
MDR Text Key5489656
Report Number1828100-2015-00132
Device Sequence Number1
Product Code DTY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K902654
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number100
Device Catalogue Number100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/06/2015
Initial Date FDA Received02/10/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-