• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES HAGENDORF T-PAL TRIAL SPACER 10MM X 28MM 10MM HEIGHT; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES HAGENDORF T-PAL TRIAL SPACER 10MM X 28MM 10MM HEIGHT; TEMPLATE Back to Search Results
Catalog Number 03.812.310
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Device report from synthes (b)(6) reports an event in (b)(6) follows: it was reported that trial and implant turned prematurely in the disc space whilst inserter was in the locked/closed position.When we checked the instrument outside the patient you could push the implant and trial over into turned position with very little force.There was no patient harm.It was reported that the surgery was prolonged about 30 minutes.This is report 2 of 4 for (b)(4).
 
Manufacturer Narrative
Device is an instrument and is not implanted/explanted.The investigation could not be completed; no conclusion could be drawn, as no product was received.A device history review was conducted.The report indicates that no ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-PAL TRIAL SPACER 10MM X 28MM 10MM HEIGHT
Type of Device
TEMPLATE
Manufacturer (Section D)
SYNTHES HAGENDORF
im bifang 6
hagendorf CH46 14
SZ  CH4614
Manufacturer (Section G)
SYNTHES HAGENDORF
im bifang 6
hagendorf CH46 14
SZ   CH4614
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4512257
MDR Text Key5409635
Report Number3003875359-2015-10087
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.812.310
Device Lot Number7600871
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/09/2015
Initial Date FDA Received02/12/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-